ARTHROCARE CORP. OPUS SPEEDSCREW IMPLANT; SCREW, FIXATION, BONE Back to Search Results
Model Number OM-6500
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an unknown procedure, the tip of the opus speedscrew implant broke during impact.All parts of the broken screw were retrieved from the patient.The procedure was completed using a back-up device.It is unknown if there was a delay.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A review of the material specifications found that tensile strength should be verified to meet specifications, and a certificate of material analysis is required with each shipment.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports, the tip of the screw broke in the patient during impact in an unknown procedure.Reportedly, all parts of the broken screw were retrieved from the patient and the procedure was completed with a backup device.It is unknown if a procedural delay occurred, and no further complications were reported.Smith and nephew has not received the requested patient specific documentation, the explanted device, and/or adequate materials to fully evaluate the root cause of the complaint.Patient impact beyond that which was reported (tip of the screw broke in patient during impact) could not be determined.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPUS SPEEDSCREW IMPLANT
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14089737
Report Number3006524618-2022-00198
Device Sequence Number17370387
Product Code HWC
UDI-Device Identifier00817470005196
UDI-Public00817470005196
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K081893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/06/2024
Device Model NumberOM-6500
Device Catalogue NumberOM-6500
Device Lot Number2075333
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer11/18/2022
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date11/21/2022
Date Device Manufactured07/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-