• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CETERIX ORTHOPAEDICS, INC NOVOSTITCH PRO MENISCAL RPR SYS 0; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CETERIX ORTHOPAEDICS, INC NOVOSTITCH PRO MENISCAL RPR SYS 0; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number CTX-A004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a meniscus repair, the axis for moving the upper jaw of the novostitch fractured in the body, it could not be manipulated.The broken piece could not be removed from the patient.It is unknown how the procedure was completed.A non-significant delay was reported.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states the externally communicating device is neither manufactured nor intended for implantation; and is not approved for long-term internal tissue exposure and long-term implantation data is not available.A photo of the of the device was reviewed and supports the reported breakage.Although unlikely, the potential for corrosion and local irritation/migration of the device fragments could not be ruled out.The root cause of the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Additional Manufacturer Narrative
Corrected data: d3 & g1 (manufacturing registered site).Fei number associated to this report should be (b)(4) instead of (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOSTITCH PRO MENISCAL RPR SYS 0
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
CETERIX ORTHOPAEDICS, INC
6500 kaiser dr ste 120
fremont CA 94555
Manufacturer (Section G)
CETERIX ORTHOPAEDICS, INC
6500 kaiser dr ste 120
fremont CA 94555
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14089958
Report Number3006524618-2022-00199
Device Sequence Number11978225
Product Code GAT
UDI-Device Identifier00853541006051
UDI-Public00853541006051
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181772
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/05/2023
Device Model NumberCTX-A004
Device Catalogue NumberCTX-A004
Device Lot NumberM210192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer11/02/2023
05/06/2025
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date11/08/2023
05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-