Model Number IJS-E Base Plate Assembly
Medical Device Problem Code Installation-Related Problem (2965)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
Surgeon identified loose ijs construct 10 days post implantation.
 
Additional Manufacturer Narrative
No device problems were identified.Revision procedure was completed successfully and no further issues have been identified.The ifu states "if the locking screws or the axis pin is not fully tightened, the construct may loosen, shift, and/or disassemble." as per witness, it was stated that the surgeon did not properly tighten the construct which contributed to malfunction post-implantation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERNAL JOINT STABILIZER - ELBOW BASE PLATE ASSEMBLY
Common Device Name
IJS- ELBOW ASSEMBLY
Manufacturer (Section D)
SKELETAL DYNAMICS
7300 n kendall drive suite 400
miami FL 33156
Manufacturer Contact
7300 n kendall drive
miami, FL 33156
MDR Report Key14091181
Report Number3006742481-2022-00003
Device Sequence Number20281791
Product Code OZI
UDI-Device Identifier00841506104904
UDI-Public00841506104904
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K153208
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Remedial Action Patient Monitoring
Type of Report Initial
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIJS-E Base Plate Assembly
Device Catalogue NumberIJS-ELB-BPA
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
Patient SexMale
Patient Weight88 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-