Investigations: visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve operates within the accepted tolerance in the horizontal position.Internal inspection: after dismantling of the valve, deposits were found in progav 2.0.Result: based on our investigation, we are unable to substantiate the clinical claim of under-drainage.At the time of our investigation, the opening pressure of the valve was within the specified tolerances.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
|
It was reported that a progav 2.0 (part # fx410t) was implanted during a procedure performed on january 18, 2022.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), height: 71 centimeters (cm), weight: (b)(6), gender: male.Same event as medwatch mdr id 3004721439-2022-00077.
|