• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. HYDROFRAME-10-AV; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. HYDROFRAME-10-AV; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 100930HFIL-V-A2
Medical Device Problem Codes Detachment of Device or Device Component (2907); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an embolization treatment of a ruptured giant acom aneurysm (12mm), the second coil was a hydrofill 9x30.The coil was pushed through a prowler 14 microcatheter.Approximately 50% of the coil was inside the aneurysm and resistance was encountered.It was decided to withdraw the coil.During the withdraw maneuver, the coil detached from the delivery pusher partially within the aneurysm and the microcatheter.The coil was successfully removed using a snare over the transfemoral sheath.Additional coils were place with good result.The patient was reported to be intubated on icu.
 
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was discarded by user facility and not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed at this time.The instructions for use (ifu) identifies premature coil detachment as potential complication a associated with use of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROFRAME-10-AV
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key14094804
Report Number2032493-2022-00153
Device Sequence Number20248182
Product Code HCG
UDI-Device Identifier00816777023315
UDI-Public(01)00816777023315(11)201106(17)251031(10)2011061W5
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K090357
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model Number100930HFIL-V-A2
Device Catalogue Number100930HFIL-V
Device Lot Number2011061W5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Initial Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-