MAUDE Adverse Event Report: BIOMET 3I TITANIUM HEXED UNISCREW; DENTAL SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I TITANIUM HEXED UNISCREW; DENTAL SCREW Back to Search Results
Model Number UNIHT
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
Doctor reported that screw fractured at the thread of the screw.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient identifier unknown / not provided.Age at time of the event unknown / not provided.Gender unknown / not provided.Weight unknown / not provided.Additional device information unknown / not provided.Device manufacturer date unknown / not provided.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Additional report submitted to updated udi number that has not been included in the previous reports.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The following sections have been updated: b4: date of this report b5: describe event or problem d4: expiration date d4: udi unknown g3: date received by the manufacturer h1: type of report, follow-up number h2: follow up type h3: device evaluated by manufacturer: change ¿no' to 'yes' *h4: device manufacture date h6: evaluation codes h10: additional narrative one screw was returned for investigation.Visual inspection via naked eye and camera magnification confirmed that the screw was fractured at the threads.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.Pre-existing conditions noted on the complaint: bruxism; patient has night-guard.The reported device was located on unknown tooth site.Pictures or x-ray images were not provided.Review of appropriate documentation: documents reviewed: biomet 3i restorative products ifu (p-iis086gr) rev f - october 2019.Information identified: 'precautions' 'breakage' 'warning' per the applicable ifu, breakage may occur when device is loaded beyond its functional capability.Dhr review: dhr review was completed for the subject lot number (1210940).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: a lot specific complaint history review was conducted for the subject lot number (1210940) and revealed that there are no existing nonconformances/capa/hhe/d/ie/product holds for the reported device related to the reported event for similar events (complaint category keywords: fracture screw).Post market trend review: july post market trending was reviewed and there were no negative trends or corrective actions for the reported event (fracture screw) and device (uniht).Based on the available information, device malfunction did occur and the reported event was confirmed following inspection and evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITANIUM HEXED UNISCREW
Common Device Name
DENTAL SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14098941
Report Number0001038806-2022-00517
Device Sequence Number4676727
Product Code NHA
UDI-Device Identifier00844868017318
UDI-Public(01)00844868017318(17)221019(10)1210940
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K072642
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/19/2022
Device Model NumberUNIHT
Device Catalogue NumberUNIHT
Device Lot Number1210940
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer09/13/2022
09/27/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date09/14/2022
09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-