| Model Number |
SD980.108 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
02/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Mandibular implant has loosened and needs to be fixated again.
|
| |
|
Additional Manufacturer Narrative
|
|
Device met specifications, screws not manufactured by materialise.Root cause is identified as the patient condition.Device still implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device met specifications, screws not manufactured by materialise.Root cause is identified as the patient condition.Surgeon believes the patient has an undiagnosed bone disease which is leading to impaired bone healing.Replacement surgery is planned for october 2022 (d6b).
|
| |
|
Event or Problem Description
|
|
Both mandibular implants have loosened and will need to be replaced by new implants.
|
| |
|