Model Number SD980.108
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 02/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Mandibular implant has loosened and needs to be fixated again.
 
Additional Manufacturer Narrative
Device met specifications, screws not manufactured by materialise.Root cause is identified as the patient condition.Device still implanted.
 
Additional Manufacturer Narrative
Device met specifications, screws not manufactured by materialise.Root cause is identified as the patient condition.Surgeon believes the patient has an undiagnosed bone disease which is leading to impaired bone healing.Replacement surgery is planned for october 2022 (d6b).
 
Event or Problem Description
Both mandibular implants have loosened and will need to be replaced by new implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUMATCH
Common Device Name
TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED IMPLANT FOR MANDIBLE, MEDIUM
Manufacturer (Section D)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE  3001
Manufacturer (Section G)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE   3001
Manufacturer Contact
jenny jones
technologielaan 15
leuven, vlaams-brabant 3001
BE   3001
MDR Report Key14099138
Report Number3003998208-2022-00008
Device Sequence Number12355021
Product Code JEY
UDI-Device Identifier05420060381089
UDI-Public05420060381089
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K173039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSD980.108
Device Catalogue NumberSD980.108
Device Lot NumberME21HORMOH
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
Patient SexFemale
-
-