ZIMMER DENTAL ABUT TAPERED ONE-PIECE SCR-V 3.5MM 2MM; DENTAL ABUTMENT Back to Search Results
Model Number TAC2
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Age at time of the event unknown / not provided.Weight unknown / not provided.Additional device information unknown / not provided.Premarket identification k013227.Device manufacturer date unknown / not provided.
 
Event or Problem Description
It was reported that the abutment screws fractured at tooth location 46.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).One abut tapered one-piece scr-v 3.5mm 2mm, tac2 was returned for investigation.Visual inspection via naked eye and camera magnification was completed.Fracture was identified at the screw threads of the abutment.Device history record (dhr) review was completed for the subject lot number (64090522).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (64090522) for similar event (complaint category keywords: fracture screw) and revealed no existing non-conformances/capa/hhe/d/ie/product holds for the reported product.Based on the available information, device malfunction did occur and the reported event was confirmed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABUT TAPERED ONE-PIECE SCR-V 3.5MM 2MM
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14101314
Report Number0002023141-2022-00904
Device Sequence Number20404533
Product Code DZE
UDI-Device Identifier00889024018419
UDI-Public(01)00889024018419(17)230630(10)64090522
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K011028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2023
Device Model NumberTAC2
Device Catalogue NumberTAC2
Device Lot Number64090522
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-