|
Us legal: it was reported that, after a right bhr surgery was performed on (b)(6) 2011 due to osteoarthritis, severe pain and limitation of motion and function, the patient experienced symptoms that raised suspicion of elevated cobalt levels.This adverse event was treated with a revision surgery on (b)(6) 2019 in which minimal synovitis around the joint was observed.Current patient¿s health status is unknown.
|
|
It was reported that a right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the acetabular cup and femoral head was performed using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.No other similar complaints were identified to involve this batch for the acetabular cup and femoral head.No other similar complaints were identified for the part number and the reported failure mode for the femoral head.Other similar complaints were identified for the part number and the reported failure mode for the acetabular cup.This will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.The available medical documents were reviewed.With the information provided the clinical root cause of the reported pain, elevated cobalt and minimal synovitis cannot be confirmed.It cannot be concluded the reported events/clinical reactions were associated with a mal performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
|