Manufacturer's investigation conclusion: a direct correlation between heartmate 3 lvas, serial number (b)(6), and the reported events could not be conclusively established through this evaluation.The account submitted log files for review.The system controller event log file contains data from (b)(6) 2022 at 21:07:41 through 18:22:23 and (b)(6) 2022 at 14:42:50 through 17:15:39.The driveline was disconnected on (b)(6) 2022 at 14:35:24 and reconnected at 14:48:29, again disconnected at 14:50:29 during a controller exchange.The controller was shut off on (b)(6) 2022 at 15:03:48.Prior to the driveline being disconnected, multiple backup battery faults were captured.Transient pi events were observed throughout the log file, resulting in momentary decreases in speed per design.The pump appeared to function as intended.The heartmate 3 lvad, serial number (b)(6), was not returned for evaluation.The heartmate 3 lvas ifu, rev.C, is currently available.This ifu death as an adverse event that may be associated with the use of the heartmate 3 left ventricular assist system.In section 2 "system operations" (under "system controller warnings and cautions"), the ifu states, "if the driveline disconnects from the system controller, the pump stops.If the driveline disconnects from the system controller, promptly reconnect it to resume pump operation." section 2 also contains a section titled "connecting the driveline to the system controller", which provides instructions on connecting/disconnecting the driveline to/from the system controller and making sure that it is fully and properly inserted into the system controller socket.Section 7 "alarms and troubleshooting" outlines all system controller alarms, including the driveline disconnected hazard alarm, as well as how to respond to each alarm condition.Section 6 (under "educating and training patients, families, and caregivers") explains that during the patient selection, preimplant, and postoperative period, the patient must receive instructions regarding the operation and care of every system component (including the driveline and what to do in an emergency).The heartmate 3 lvas patient handbook, rev.C, warns "the pump will stop if the driveline is disconnected from the system controller.If the driveline disconnects from the system controller, reconnect it right away to restart the pump.The pump cannot run without power." in addition, the patient handbook states, "call your hospital contact if you think that, for any reason, any portion of your equipment is not functioning as usual, is broken, or you are uncomfortable with the operation of the equipment." the patient handbook also contains a section on handling emergencies.Section 5 ¿alarms and troubleshooting¿ contains information regarding all system controller alarms, including the driveline disconnected hazard alarm, and the proper actions associated with the relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The relevant sections of the device history records for the percutaneous lead were also reviewed and showed no deviations from manufacturing or quality assurance specifications.The implant kit was shipped on (b)(6) 2018.No further information was provided.The manufacturer is closing the file on this event.
|