INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG Back to Search Results
Model Number 82-3832
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code Cerebrospinal Fluid Leakage (1772)
Date of Event 03/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a hakim valve (id 823832) was implanted on (b)(6) 2022.The valve was with pressure 40 mmhg and could not reach the expected pressure setting 90mmhg.On (b)(6) 2022, during the exploratory surgery for shunt tube adjustment, the physician found that the valve has a crack and a minimal csf leaking.Therefore, the physician changed with a new id 823832 valve.The patient was stable after the revision procedure.
 
Additional Manufacturer Narrative
The hakim valve (id 823832) was not returned for evaluation (according to customer in china: due to the covid-19 quarantine, all the products cannot be returned at current stage).Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.However, a probable root cause for the reported complaint is due to user¿s error.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The hakim valve was returned for evaluation: failure analysis - the position of the cam when valve was received was 70mmh2o.The valve was visually inspected; the silicone housing at the point where the connector should be is damaged crushed/torn (possible due to a hard knock), marks were also noted on the connector.The complaint is confirmed.The root cause for the issue reported by the customer could be due to a sharp or pointed object coming into contact with the silicone housing, as noted in the ifu, silicone has a low cut/tear resistance.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Common Device Name
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14102746
Report Number3013886523-2022-00156
Device Sequence Number12355046
Product Code JXG
UDI-Device Identifier10886704041825
UDI-Public(01)10886704041825(17)250331(10)4660678
Combination Product (Y/N)N
PMA/510(K) Number
K053107
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number82-3832
Device Catalogue Number823832
Device Lot Number4660678
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer04/14/2022
08/26/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date04/19/2022
09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-