PHILIPS ULTRASOUND SPARQ ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 989605394911
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The initial review of the suspect ultrasound system¿s data logs did not indicate a hardware failure.A philips service engineer is scheduled to visit the customer site to troubleshoot the cause of the reported failure.Results from the event evaluation and system repair will be included in a follow up report upon investigation completion.
 
Event or Problem Description
A customer reported their sparq ultrasound system locked up and displayed vertical line artifacts during transesophageal transducer use for exploratory surgery.The procedure was able to continue without the sparq system using the images previously acquired.There was no patient injury associated with this event.
 
Additional Manufacturer Narrative
A thorough investigation was performed to identify the root cause of the reported issue, including a technical and workflow evaluation of this event.The engineering team was unable to reproduce the issue as described by the customer.The system is currently in use at the customer site with no additional similar issues reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPARQ ULTRASOUND SYSTEM
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key14103428
Report Number3019216-2022-00046
Device Sequence Number4676862
Product Code IYN
UDI-Device Identifier00884838028098
UDI-Public00884838028098
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K091804
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number989605394911
Device Catalogue Number795090
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer03/30/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/25/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-