ARROW INTERNATIONAL LLC INTRA-AORTIC BALLOON CATHETER KIT; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN917531
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Event Description
It was reported that "the intraaortic balloon pump broke." it was stated that blood was noted in the line.As a result, therapy was discontinued.There was no report of patient complications, serious injury or death.
 
Manufacturer Narrative
(b)(4).There is an investigation ongoing into this reported complaint.A supplement follow-up mdr report will be submitted with the investigation results upon completion of the investigation.
 
Manufacturer Narrative
Qn# (b)(4).The product was not returned for investigation.The reported complaint of iab blood in helium pathway is not able to be confirmed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event Description
It was reported that "the intraaortic ballon pump broke." it was stated that blood was noted in the line.As a result, therapy was discontinued.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRA-AORTIC BALLOON CATHETER KIT
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key14104135
MDR Text Key289202687
Report Number3010532612-2022-00128
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902034724
UDI-Public00801902034724
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K021462
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberIPN917531
Device Catalogue NumberIAB-05830-LWS
Device Lot Number18F21K0041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received04/13/2022
Supplement Dates Manufacturer Received05/03/2022
Supplement Dates FDA Received05/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
-
-