• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number RA002-4545SLR70
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod would not distract.No patient adverse event was reported.
 
Additional Manufacturer Narrative
Complaint 1 of 2.Visual inspection of the returned magec rod revealed that the rod was partially distracted with some wear/score marks on the distraction rod consistent with expected markings due to incremental distraction.The wear/score marks observed on the distraction rod would indicate it was extended approximately 2-3 mm from its initial position.X-ray images of the internal components showed a broken radial ball bearing.The rod was functionally tested; during testing the rod was unable to distract or retract with the manual distractor as well as the external remote controller (erc).The rod was therefore determined to be jammed.Sectioning of the rod determined that the distraction rod could not be separated from the housing tube without the use of high force, which indicates the distraction rod was out of alignment.It is possible, bending forces applied to the rod from patient anatomy/activity may have caused the distraction rod to become wedged into the housing tube/become out of alignment, which may be a cause for the broken radial ball bearing.The rod was determined to be out of alignment possibly due to excessive axial loading.Additionally, it is possible that patient activities may be lined to the reported failure.Per the product eifu, contraindications include: "patients and/or families unwilling or incapable of following postoperative care instructions." review of the device history records revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
 
Event or Problem Description
No additional information was provided.
 
Event or Problem Description
Additional information was received.
 
Additional Manufacturer Narrative
Corrected data: d4.Additional data: b5, e1, g3, g6, h2, h6, h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14104317
Report Number3006179046-2022-00166
Device Sequence Number14649694
Product Code PGN
UDI-Device Identifier00856719002145
UDI-Public856719002145
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K201543
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRA002-4545SLR70
Device Lot Number150810-04-1AF
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer05/03/2022
06/20/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date05/16/2022
06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-