• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-80D265
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
Information was received that a revision procedure was performed due to the nail breaking.No patient adverse event was reported.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional data: b5, d9, h3, h6, h10 corrected data: d4 (model number and udi number) device evaluation: upon return, visual inspection of the returned stryde nail revealed that the housing tube was broken.Based on visual inspection the reported failure can be confirmed.Examining the broken nail closely, it was determined that the broken ends were stretched not straight which is a sign that the nail was subjected to an excessive bending force which may have been generated from weight-bearing activities.The work order was reviewed and confirmed the device passed all inspections and acceptance tests prior to shipment.No material/manufacturing issue was found.Per the precice stryde instructions for use, ¿the precice stryde¿ nail cannot withstand the stresses of full weight-bearing.Patients should utilize external support and/or restrict activities as directed by the physician until consolidation occurs¿.In addition, the reported nail was implanted for more than a year which exceeds the guidelines with the ifu.Device records review: a review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.
 
Event or Problem Description
No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14104414
Report Number3006179046-2022-00169
Device Sequence Number6427412
Product Code HSB
UDI-Device Identifier00887517951182
UDI-Public887517951182
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K180503
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPS10.0-80D265
Device Lot Number0061709AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-