MAUDE Adverse Event Report: ENCORE MEDICAL L.P TAPERFILL HIP; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 12

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P TAPERFILL HIP; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 12 Back to Search Results
Model Number 425-97-012
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 03/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Complaint - surgeon trialed with the broach size that was appropriate for the patient.When the implant was placed it subsided causing the hip to be unstable.Implants had to be taken out and replaced.
 
Additional Manufacturer Narrative
The reason for this complaint was reported as broaching technique causing the stem to subside.The healthcare professional indicated this event occurred during treatment near the patient.There was another suitable device available, however, the incident did cause a 40 minute delay in surgery.The surgery was completed as intended, there was no risk or adverse event reported and the item was inspected prior to surgery and was found to be acceptable.The device was not returned to manufacturer and evaluated by registered medical assistant (rma) at djo surgical.A review of the device history record (dhr) revealed, when released for use, the items met design and manufacturing requirements.There were no ncmrs associated with the production of this instrument.Complaint database review showed previous complaints but there were no indications that this component has a design or material deficiency.To date number of complaints against this lot no.: 0 summary of complaints to date: finished goods.S103 - fit 1 finished goods.S301 - taper damage 1 finished goods.S800 - revision surgery 3 finished goods.S803 - dislocation 1 finished goods.S807 - pain 1 finished goods.S808 - infection 3 finished goods.S809 - trauma 6 finished goods.S810 - device loosening 2 this event is possibly attributable to broaching technique causing the stem to subside.This is not an event associated with a product failure, malfunction or issue.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERFILL HIP
Common Device Name
TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 12
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14105056
Report Number1644408-2022-00504
Device Sequence Number20439981
Product Code LPH
UDI-Device Identifier00888912078702
UDI-Public(01)00888912078702
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number425-97-012
Device Catalogue Number425-97-012
Device Lot Number177R2237
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date04/13/2022
Supplement Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age67 YR
Patient SexFemale
-
-