• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES(IRELAND) CUSA CLARITY 23 KHZ HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES(IRELAND) CUSA CLARITY 23 KHZ HANDPIECE Back to Search Results
Model Number C7023
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cusa clarity handpiece was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies hat could be associated with the complaint incident was observed.Reported failure of temperature issue during sterilization was unconfirmed.The device was manufactured in (2020).Review of service history confirmed this is the 1st service.Corrective action was unable to be implemented as the device was not returned for evaluation.Based on the reported failure of "temperature issue during sterilization", further information was requested.No response to questions was received.There is insufficient information available to determine exact root cause but based on report provided by customer, the methodology used to sterilize the device deviates from that stipulated in the device ifu.However, a complete investigation into root cause is unable to be completed at this time.Should the device be returned for analysis the complaint will be reopened and an investigation will be completed.In the interim, we have recommended the following: customer to provide answer to questions requested to assist understanding of issue and sterilization methodology used.Customer to reference device ifu and reference sections relating to sterilization procedure.At present, we consider this complaint to be closed.
 
Event or Problem Description
This is 1 of 2 reports linked to mfg report number 3006697299-2022-00053: a facility reported that during revalidation of autoclaves, the cusa clarity 23 khz handpiece (c7023) were tested, and 7 tests on the same instrument showed that the temperature in the instrument's lumen did not reach the correct temperature for sterilization (134 degrees for 4 minutes) in relation to danish and eu standards.The temperature only reached132 degrees as maximum temperature in a very short time.The device was already used on patient was this was discovered, thus there was no patient injury or increased surgery time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSA CLARITY 23 KHZ HANDPIECE
Common Device Name
CUSA CLARITY
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
sragh, tullamore, co.offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14105188
Report Number3006697299-2022-00054
Device Sequence Number20002589
Product Code LFL
UDI-Device Identifier10381780466314
UDI-Public(01)10381780466314
Combination Product (Y/N)N
PMA/510(K) Number
K182809
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC7023
Device Catalogue NumberC7023
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Initial Report FDA Received Date04/13/2022
Date Device Manufactured08/05/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-