Model Number BTD05
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unk.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that btd05 devices was returned with the dissector tip separated.In addition, the dissector tip of the device was returned inside a plastic bag with.Upon evaluation, the dissector tip was found separated from the shaft, evidence of adhesive was found in the tip of the shaft as the cotton appears to have been pulled off.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached regarding what may have caused the reported incident.The manufacturing records couldn't be reviewed as the batch number is unknown.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a laparoscopic anterior resection, it was found that the cotton tip had fallen off when the device was removed from the patient.The procedure was interrupted to find it.It was retrieved.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ENDOPATH*BLUNT TIP DISSECTOR
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
*  
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
MDR Report Key14106390
Report Number3005075853-2022-02361
Device Sequence Number12369018
Product Code MDM
UDI-Device Identifier10705036011873
UDI-Public10705036011873
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBTD05
Device Catalogue NumberBTD05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Initial Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1