Model Number HST III SEAL (4.5MM), 5-PACK |
Device Problems
Failure to Unfold or Unwrap (1669); Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/17/2022 |
Event Type
malfunction
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hst iii seal (4.5mm) seal had fallen off when the delivery system was pulled out of the loading device.The seal was not loaded properly and could not be unfolded.The operation was completed with the new hs #.The patient had no health hazard.
|
|
Manufacturer Narrative
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Trackwise # (b)(4).Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213/67) the review of the historical data indicates that no other similar complaints was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period apr -2020 through mar -2022 was reviewed.There were no triggers identified for the review period communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
|
|