BAYLIS MEDICAL COMPANY INC. VERSACROSS RF WIRE; CATHETER, SEPTOSTOMY Back to Search Results
Model Number VXW-35-180-P
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
A 20-minute procedural delay was reported due to knotting of the versacross rf wire (rf wire) while attempting to pull the wire into the versacross transseptal sheath for the drop down of the transseptal puncture workflow.The versacross transseptal sheath became kinked in the process.The procedure was able to continue after a new versacross rf wire and versacross transseptal sheath were used to perform a successful transseptal puncture.While there was no patient injury reported, baylis medical company inc.Has decided to report this event due to the procedural delay.
 
Additional Manufacturer Narrative
While there was no patient injury reported, baylis medical company inc.Has decided to report this event due to the 20-minute procedural delay.Dhr review was completed for the lot in question, and it was confirmed that all devices met relevant requirements prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS RF WIRE
Common Device Name
CATHETER, SEPTOSTOMY
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal, quebec H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
2775 matheson blvd. east
mississauga, ontario L4W 4 P7
CA   L4W 4P7
Manufacturer Contact
meghal khakhar
2775 matheson blvd. east
mississauga, ontario L4W 4-P7
CA   L4W 4P7
MDR Report Key14106901
Report Number9710452-2022-00026
Device Sequence Number17340745
Product Code DXF
UDI-Device Identifier00685447003298
UDI-Public(01)00685447003298(17)241028(10)VXFA271021
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K150709
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2024
Device Model NumberVXW-35-180-P
Device Lot NumberVXFA271021
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Initial Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-