| Model Number |
PED2-375-12 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that a ped2 pipeline failed to open in the distal section.Doctor thinks it was because the rist catheter was not giving much support due to anatomy (femoral).The patient was being treated for an unruptured, saccular aneurysm in the ophthalmic location.The max diameter was 4mm and the neck diameter was 2.5mm.The distal landing zone was 3.3mm and the proximal landing zone was 3.7mm.Vessel tortuosity was moderate.Dual antiplatelet treatment was administered.The pru level was 166.Angiographic result post procedure was good.The devices were prepared as indicated in the instructions for use (ifu).The pipeline was used for an indication that is not approved.(watson label). the pipeline was positioned in a bend.More than 50% of the pipeline had been deployed when it failed to open.The pipeline was resheathed more than 2 ties.The pipeline was removed from the patient with the microcatheter.No symptoms were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported there was no catheter kickback.
|
| |
|