Model Number N/A
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a patient required placement of a nester platinum embolization microcoil for embolization of gastric varices using an echoendoscope.The coil was pushed through the introducer on the proximal end of an endoscopic ultrasound fine needle until it was fully inside the needle, and then the stylet needle was used to push the coil.However, the operator could not advance the coil into the patient.The physician removed the device and attempted again with another coil of the same lot.The same difficulty was experienced.The procedure was unable to be completed.The physician decided to continue watching the patient and may attempt to complete the procedure at a later date.At this time, there is no confirmation of a planned additional procedure.Images provided from the customer show both coils partially exited the tip of the needle.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: sutter cpmc van ness campus (united states) informed cook on 23mar2022 of an incident involving a patient requiring the placement of a nester platinum embolization microcoil (rpn: mwce-18-14-10-nester-01; lot: 14207549) for embolization of gastric varices using an echoendoscope.The coil was pushed through the introducer on the proximal end of an endoscopic ultrasound fine needle until it was fully inside the needle, and then the stylet needle was used to push the coil.However, the operator could not advance the coil into the patient.The physician removed the device and attempted again with another coil of the same lot.The same difficulty was experienced.The procedure was unable to be completed.The physician decided to continue watching the patient and may attempt to complete the procedure at a later date.At this time, there is no confirmation of a planned additional procedure.Images provided from the customer show both coils partially exited the tip of the needle.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), and instructions for use (ifu), as well as a visual inspection of the returned devices, were conducted during the investigation.The complaint devices were not returned; however, the customer did return seven unused, unopened devices from the same lot for evaluation.All seven devices returned were opened, and there was no trouble getting the coil to push out from the loading cartridge using the pusher stylet.The failure mode was unable to be confirmed.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot 14207549 revealed no related non-conformances.A database search for complaints on the reported lot found no additional complaints reported from the field.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in field.Cook also reviewed product labeling.The ifu [t_nec2_rev0] packaged with the device contains the following in relation to the reported failure mode: "instructions for use: for.018 inch microcoils: 1.Perform an angiogram prior to embolization to determine optimal catheter position.2.Push the loading cartridge completely into the delivery catheter.3.Advanced the luer lock connector fitting toward the catheter hub and lock into place.4.Use the pusher stylet to load the nestor embolization microcoil into the delivery catheter.Note: the stylet must be pushed as far as possible into the loading cartridge to ensure proper loading.Remove the pusher stylet and loading cartridge.5.To obtain secure placement of the nester embolization microcoil., we recommend the use of the pusher stylet to push the microcoil further into the delivery catheter.6.Once coil is positioned into delivery catheter, verify position of the catheter tip prior to deployment.Coil design permits delivery into the target vessel by saline/contrast flush or by push technique using an appropriately sized wire guide or pusher.7.Perform final angiogram to confirm coil position within target vessel." based on the information provided, inspection of the unused, unopened returned devices, and the results of the investigation, cook has determined that failure to follow instructions contributed to this failure due to the customer attempting to advance the coil through a needle and not a delivery catheter.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NESTER PLATINUM EMBOLIZATION MICROCOIL
Common Device Name
KRD DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key14110984
Report Number1820334-2022-00569
Device Sequence Number6229709
Product Code KRD
UDI-Device Identifier00827002473256
UDI-Public(01)00827002473256(17)260926(10)14207549
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153778
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2026
Device Model NumberN/A
Device Catalogue NumberMWCE-18-14-10-NESTER-01
Device Lot Number14207549
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date01/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COOK ECHO-3-22
Patient SexUnknown
-
-