• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-017
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported on (b)(6) 2022, a 17mm amplatzer septal occluder was selected for implant.During deployment, the left disc appeared oval and was protruding into the left atrium.It was noted that two attempts were made to deploy the occluder before the device was removed and replaced with a new 17mm amplatzer septal occluder.The new device was successfully implanted without any complications.The patient was reported to be in stable condition.There was no clinically significant delay in procedure and no adverse patient effects reported.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.The cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
It was reported on (b)(6) 2022, a 17mm amplatzer septal occluder was selected for implant.During deployment, the left disc appeared oval and was protruding into the left atrium.It was noted that two attempts were made to deploy the occluder before the device was removed and replaced with a new 17mm amplatzer septal occluder.The new device was successfully implanted without any complications.The patient was reported to be in stable condition.There was no clinically significant delay in procedure and no adverse patient effects reported.Subsequent to the initially filed information, additional information was received that the delivery system was a 7f amplatzer trevisio intravascular delivery system.No angulation or kink was noted in the delivery system.There was no interaction with cardiac structures during the deployment of the first device.The patient remained hemodynamically stable throughout the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14111083
Report Number2135147-2022-00172
Device Sequence Number2121591
Product Code MLV
UDI-Device Identifier05415067019257
UDI-Public05415067019257
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model Number9-ASD-017
Device Catalogue Number9-ASD-017
Device Lot Number8076251
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer05/19/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-