Model Number NX1000-3
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Heart Problem (4454); Swelling/ Edema (4577)
Event Date 03/31/2022
Event Type  Injury  
Manufacturer Narrative
A review of the device history record (dhr) was conducted which confirmed the device met all quality criteria and manufacturing specifications prior to release. Factors outside the scope of nxstage therapy can impact the patient's weight, these include but are not limited to the accuracy with which intake is recorded, the weighing techniques used, and the patient's comorbidities. The nxstage system one user guide provides information on the risks associated with dialysis therapy and suggested actions, instructions, cautions and warnings to enable patients to treat safely and effectively. Udi: (b)(4).
 
Event Description
A report was received on (b)(6) 2022 from the home therapy nurse (htn) of a (b)(6) female patient with a medical history including end stage renal disease, who stated an insufficient amount of programmed fluid was removed during home hemodialysis treatments, most recently on (b)(6) 2022. Per the htn, the patient has been progressively fluid overloaded since late on (b)(6) 2021 without a differential causal diagnosis provided. Additional information was received on 01 apr 2022 from the (htn) who stated the patient has developed cardiac issues (nos) related to fluid overload, has had pitting edema on her lower legs, possible abdominal fluid retention, with a planned procedure to remove thoracic fluid buildup (thoracentesis). Although the cause of the fluid overload has not been determined, a possible relationship to the device cannot be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNXSTAGE SYSTEM ONE
Type of DeviceHIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX 22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
MDR Report Key14111852
MDR Text Key289316898
Report Number3003464075-2022-00010
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K122051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation
Type of Report Initial
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberNX1000-3
Device Catalogue NumberCHRONIC HI-FLOW CYCLER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/14/2022 Patient Sequence Number: 1
-
-