• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO2 CATHETER WITH SHERLOCK 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO2 CATHETER WITH SHERLOCK 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by healthcare professional "i have a catheter that i pulled today that was damaged.The picc was leaking significantly when it was flushed.The pt reports that this did not happen the day before.I immediately removed the line and there is a horizontal crack at the 6 cm mark.This [picc placement] would not have been in any area with a significant bend such as the arm pit.".
 
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of (b)(4) showed no other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a catheter leak was confirmed.The product returned for evaluation was one 4fr sl powerpicc solo catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed between 5 and 6 cm mark.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: damage which was circumferentially aligned ¿ fracture edges which were rounded and polished due to repeated material wear ¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together ¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing) an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.H3 other text: evaluation findings are in section h11.
 
Event or Problem Description
It was reported by healthcare professional "i have a catheter that i pulled today that was damaged.The picc was leaking significantly when it was flushed.The pt reports that this did not happen the day before.I immediately removed the line and there is a horizontal crack at the 6 cm mark.This [picc placement] would not have been in any area with a significant bend such as the arm pit." additional info.Received: the leak was discovered prior to an infusion when the rn was flushing the line.Yes the catheter was removed and a new one was placed a bd picc stat lock was used to secure picc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO2 CATHETER WITH SHERLOCK 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14111899
Report Number3006260740-2022-01263
Device Sequence Number5276236
Product Code LJS
UDI-Device Identifier00801741154881
UDI-Public(01)00801741154881
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number1194108D2
Device Lot NumberREFX1724
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
Patient SexUnknown
-
-