DATASCOPE CORP. - FAIRFIELD UNKNOWN INTRA-AORTIC BALLOON; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number UNKNOWN
Device Problems Leak/Splash (1354); Migration (4003)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/27/2022
Event Type  Injury  
Event Description
It was reported that during intra-aortic balloon (iab) therapy blood was seen in the inner lumen line running to the flush bag.The customer proceeded to flush the line for a few minutes until it cleared.Directly after that, the patient developed some chest tightness and shortness of breath for a period of about 10 minutes.The customer wanted to know what it may mean clinically and if the iab may need to be removed from the patient.The getinge representative asked if the pump was put in standby, as per the ifu during flushing.The customer stated that they had not put the pump in standby.The getinge representative reviewed the importance of performing flushing while in standby and the possible clinical results if not followed correctly.The customer then stated that the patient seem clinically stable and therapy was being delivered as expected.The customer ordered a chest x-ray which indicated that the iab was not in the proper position.It was later reported that the patient was stable.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN INTRA-AORTIC BALLOON
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key14112176
MDR Text Key289270574
Report Number2248146-2022-00300
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
-
-