• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number PNDL, 3BV TVH TP 31G 8MMTC 100CT 30/CS BLUE
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported complaint for inaccurate dispense/aspiration of the 31g pen needle.Customer advised that when she attempts to give herself the insulin, the insulin in not aspirating and dispensing through the pen needle.The customer was not using compatible product (lispro injection kwik pen).The customer has been using the product out of this package for a few weeks.The customer feels well and did not report any symptoms.No medical attention associated with the use of the product was reported.
 
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on (b)(6) 2022 to ensure that the initial concern was resolved - able to establish contact with customer who stated she is still using the true plus pen needles.Customer was again advised that the pen needles are not compatible with her pen.
 
Additional Manufacturer Narrative
Sections with additional information as of 16-jun-2022: h6: updated fda¿s type, findings and conclusions codes.H10: pen needles were not returned for evaluation.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using pen needles from the same lot.No abnormalities observed on retain samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NEEDLES FG
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key14112266
Report Number1000113657-2022-00216
Device Sequence Number17669223
Product Code FMI
UDI-Device Identifier00021292012809
UDI-Public(01)00021292012809
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2026
Device Model NumberPNDL, 3BV TVH TP 31G 8MMTC 100CT 30/CS BLUE
Device Catalogue NumberS5H01D31-100TC
Device Lot Number1F620
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date03/21/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-