MAUDE Adverse Event Report: HOLOGIC, INC SELENIA MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC SELENIA MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY Back to Search Results
Model Number SEL-00007
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the system c-arm moved unconmanded.No injuries reported.A field engineer was dispatched to the site and determined that the detector rotation switch needed to be readjusted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA MAMMOGRAPHY SYSTEM
Common Device Name
FULL FIELD DIGITAL MAMMOGRAPHY
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
evelyn lopez
36 & 37 apple ridge rd
danbury, CT 06810
MDR Report Key14112281
Report Number1220984-2022-00025
Device Sequence Number2135039
Product Code MUE
UDI-Device Identifier15420045505940
UDI-Public15420045505940
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P010025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSEL-00007
Device Catalogue NumberSEL-00007
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Initial Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-