Model Number P1190A000067 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/26/2022 |
Event Type
malfunction
|
Event Description
|
Hillrom received a report from a hillrom technician stating that the head of the bed was intermittently raising by itself.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
|
|
Manufacturer Narrative
|
The hillrom technician found the right side rail cable was damaged and was shorting out.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the side rail cable assembly to resolve the issue.Based on this information, no further action is required.
|
|
Search Alerts/Recalls
|