• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC TRUE FOCUS; SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC TRUE FOCUS; SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER Back to Search Results
Model Number STRIP, TF WGN 50CTMG/DL 956970
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Note: manufacturer contacted customer in several follow-up calls to ensure the replacement products resolved the initial concern unable to establish contact with customer at this time.
 
Event or Problem Description
Consumer reported complaint for lo blood glucose test results.The customer is concerned with test results from result obtained on (b)(6) 2022 of lo.The customer¿s expected am fasting blood glucose test result range is 130-140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a back to back blood test was not performed by the customer; when the customer inserted the test strip, the low battery signal displayed on the screen and the true focus meter did not go into test mode.The battery in the true focus meter had never been changed, customer has had the true focus meter for less than one month.The product is stored according to specification in the bedroom.The test strip lot manufacturer¿s expiration date is 12/13/2023 and open vial date is (b)(6) 2022.The meter memory was not reviewed for previous test result history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE FOCUS
Common Device Name
SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key14112752
Report Number1000113657-2022-00217
Device Sequence Number20360522
Product Code NBW
UDI-Device Identifier00021292008956
UDI-Public(01)00021292008956
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K131013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/13/2023
Device Model NumberSTRIP, TF WGN 50CTMG/DL 956970
Device Lot NumberNY1014
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/21/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Initial Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-