| Model Number |
1149.0001 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The two locking caps that loosened post operatively were returned for evaluation.Visual observation shows no indentations/deformations indicative of being final tightened.The other eleven locking caps from the construct that did not loosen were also returned.Two of these do not have any indentations/deformations and the two others have extremely faint indentations indicative of being final tightened.Seven caps appear to have been torqued out properly and have the full indentation/deformation along the bottom surface.No determinations can be made as to the cause of the reported issue.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision was completed for 2 quartex threaded locking caps that loosened at c2 and c3, 2 months post operatively.
|
| |
|