Model Number G44155
Medical Device Problem Codes Break (1069); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfg: device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As initially reported, another manufacturer's atherectomy device was unable to advance through a flexor ansel guiding sheath's hub during an atherectomy procedure.Another sheath of the same type was used to successfully complete the procedure.Upon return and initial evaluation of the device, the shaft was noted to be separated below the hub and the inner coils were exposed.Additional information has been requested.
 
Event or Problem Description
No additional information has been received since the last report was submitted.
 
Additional Manufacturer Narrative
Summary of event: as initially reported, another manufacturer's atherectomy device was unable to advance through a flexor ansel guiding sheath's hub during an atherectomy procedure.Another sheath of the same type was used to successfully complete the procedure.Upon return and initial evaluation of the device, the shaft was noted to be separated below the hub and the inner coils were exposed.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The used complaint device was returned to cook for investigation.The sheath was damaged/separated.The device was received with the separated section, extending 1.4 centimeters from the white cap, attached to the hub.The coil is exposed and protruding from the point of separation.The other manufacturer¿s atherectomy device was not returned with the sheath.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu states ¿all interventional or diagnostic instruments used with this product should move freely through the valve and sheath to avoid damage.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a component failure contributed to this failure mode.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14112819
Report Number1820334-2022-00571
Device Sequence Number13987070
Product Code DYB
UDI-Device Identifier00827002441552
UDI-Public(01)00827002441552(17)241215(10)14421768
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/15/2024
Device Model NumberG44155
Device Catalogue NumberKCFW-7.0-18/38-45-RB-ANL0-HC
Device Lot Number14421768
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ROTAREX ATHERECTOMY DEVICE
Patient SexUnknown
-
-