• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH MONOJECT SAFETY NDL 25 X 5/8; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH MONOJECT SAFETY NDL 25 X 5/8; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 1182558
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the side of the orange needle had a slight split in the plastic resulting in treatment being leaked onto the nurse administering it.No harm or injury was reported.
 
Additional Manufacturer Narrative
The device history record review of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.No anomalies found on batch documentation.No samples or photos were received by the customer for evaluation.Fifty pieces of retained samples were visually checked with no anomalies found on the needle hub.Five pieces of retained samples were physically tested for stress cracking per iso guidelines and all samples passed the test.With no samples/photos of the reported issue and with no information about the handling of the device during assembling the needle with the syringe, it is not possible to identify the root cause of the issue.As a result, a corrective action is not required at this time.This complaint will be used for tracking and trending purposes and to monitor the product for reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOJECT SAFETY NDL 25 X 5/8
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
CARDINAL HEALTH
777 west st
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
777 west st
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west st
mansfield, MA 02048
5086183640
MDR Report Key14112883
Report Number1282497-2022-10686
Device Sequence Number2135137
Product Code FMI
UDI-Device Identifier10884521544710
UDI-Public10884521544710
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1182558
Device Catalogue Number1182558
Device Lot Number02102036
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date06/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-