| Model Number |
0166SI18 |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer has had 3 of the foley catheters balloons ruptured which they were using for induction.Stated that the doctor inserted the foley catheter through the cervix, and they inflated the balloon with approximately 60ml of normal saline, which did exceed the recommended amount of 30ml, but they have used them for years this way with hardly any problems.Customer also stated that the patient last week that had the foley break went for a c-section and the doctor found a part of the balloon inside the uterus.It is unknown what medical intervention was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed user related.Evaluation has been performed and observed the catheter balloon had rupture with missing pieces.A piece of missing pieces was not returned for evaluation.Based on event description, there was stated that the catheter got rupture due to user misused of the device by over inflated the catheter.The catheter was used as per not intended by the patient.A potential root cause for this failure mode could be short latex dwell time, latex dip speed out too slow causing thin sac or user related (misused of the device).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿cdc guidelines for appropriate indications for indwelling urethral catheter use ¿ patient has acute urinary retention or bladder outlet obstruction ¿ need for accurate urine output measurements ¿ use for selected surgical procedures ¿ to assist in healing of open sacral or perineal wounds ¿ patient requires prolonged immobilization ¿ to improve comfort for end of life care proper techniques for urinary catheter insertion perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in the patient record proper techniques for urinary catheter maintenance secure the foley catheter.Use the statlock® foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly using a separate, clean collection container for each patient¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer has had 3 of the foley catheters balloons ruptured which they were using for induction.Stated that the doctor inserted the foley catheter through the cervix, and they inflated the balloon with approximately 60ml of normal saline, which did exceed the recommended amount of 30ml, but they have used them for years this way with hardly any problems.Customer also stated that the patient last week that had the foley break went for a c-section and the doctor found a part of the balloon inside the uterus.It was unknown what medical intervention was provided.Per additional information received via email on 16may2022, it was reported that the customer had determined this to be user error and problem resolved.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer has had 3 of the foley catheters balloons ruptured which they were using for induction.Stated that the doctor inserted the foley catheter through the cervix, and they inflated the balloon with approximately 60ml of normal saline, which did exceed the recommended amount of 30ml, but they have used them for years this way with hardly any problems.Customer also stated that the patient last week that had the foley break went for a c-section and the doctor found a part of the balloon inside the uterus.It was unknown what medical intervention was provided.Per additional information received via email on 16may2022, it was reported that the customer had determined this to be user error and problem resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is confirmed user related.A potential root cause for this failure mode could be short latex dwell time, latex dip speed out too slow causing thin sac or user related (misused of the device).Evaluation has been performed and observed the catheter balloon had rupture with missing pieces.A piece of missing pieces was not returned for evaluation.Based on event description, there was stated that the catheter got rupture due to user misused of the device by over inflated the catheter.The catheter was used as per not intended by the patient.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: cdc guidelines for appropriate indications for indwelling.Urethral catheter use.Patient has acute urinary retention or bladder outlet obstruction.Need for accurate urine output measurements.Use for selected surgical procedures.To assist in healing of open sacral or perineal wounds patient requires prolonged immobilization.To improve comfort for end of life care.Proper techniques for urinary catheter insertion perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in the patient record proper techniques for urinary catheter maintenance secure the foley catheter.Use the statlock® foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly using a separate, clean collection container for each patient.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|