| Model Number |
1012582-19 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a perforation in the left anterior descending (lad) coronary artery.The 4.00x19 mm graftmaster covered stent was inserted into the guiding catheter and was attempted to be advanced; however, it could not cross to treat the area due to an unknown reason.The perforation was sealed with a balloon.There were no adverse patient sequela.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual and dimensional inspection was performed on the returned device.The reported failure to advance could not be tested as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported failure to advance.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a perforation in the left anterior descending (lad) coronary artery.The 4.00x19 mm graftmaster covered stent was inserted into the guiding catheter and was attempted to be advanced; however, it could not cross to treat the area due to an unknown reason.The perforation was sealed with a balloon.There were no adverse patient sequela.No additional information was provided.
|
| |
|