| Model Number |
5566-S-020 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised due to the ts stem breaking at the threaded junction of the femoral component.The femoral component and stem were revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture involving a triathlon stem was reported.The event was confirmed through clinician review of the provided medical records.Method & results: product evaluation and results: visual inspection of the returned device indicated that it has fractured.Clinician review: a review of the provided medical records by a clinical consultant indicated: these inquiries report the failure of a primary total hip replacement due to femoral component fracture necessitating revision of all three components.It also reports the failure of the revision total knee replacement due to stem extension fracture.I can confirm that these events took place since i was able to review the xrays prior to revision and after revision when the stem failed.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of femoral implant fracture and stem extension fracture are multifactorial including surgical technique factors, patient factors including activity level, weight and bmi, and implant factors.The surgeon elected to cement a press fit stem and i cannot determine if that was a contributing factor as well.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusion: an event regarding crack/fracture of the triathlon stem was reported.Visual inspection of the returned device indicated that it has fractured.A review of the provided medical records by a clinical consultant indicated: these inquiries report the failure of a primary total hip replacement due to femoral component fracture necessitating revision of all three components.It also reports the failure of the revision total knee replacement due to stem extension fracture.I can confirm that these events took place since i was able to review the xrays prior to revision and after revision when the stem failed.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of femoral implant fracture and stem extension fracture are multifactorial including surgical technique factors, patient factors including activity level, weight and bmi, and implant factors.The surgeon elected to cement a press fit stem and i cannot determine if that was a contributing factor as well.No further investigation for this event is possible at this time.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised due to the ts stem breaking at the threaded junction of the femoral component.The femoral component and stem were revised.
|
| |
|