Investigations: visual inspection the following observations were made during the visual inspection: no abnormalities detected.Permeability test: the test showed that the shuntassistant is permeable.Computer control test: the computer-controlled test showed the opening pressure of the shuntassistant, at a reference flow rate of 20 ml/h in a vertical position of the valve, to be 19,20 cmh2o.This is within the specified tolerance of 20 cmh2o +4 / -2 cmh2o.Internal inspection of product: after dismantling of the valve, deposits were found in shuntassistant.Results: based on our examination results, we can detect deposits in the outlet of the shuntassistant®.The deposits do not affect the technical properties at the time of the examination.At the time of the examination, we do not know how the above-mentioned functional impairment occurred.Proteins in the cerebrospinal fluid can temporarily influence function and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke (b)(4).From our point of view, no further regulatory actions are required.
|