• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 17MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 17MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40017
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a neurovascular surgical procedure was performed in a patient with cavernous aneurysm.During the procedure the physician felt resistance while advancing the subject flow diverter to the target lesion.At the target lesion the physician felt the subject flow diverter didn¿t feel as smooth as usual and didn¿t want to force it to open, therefore the physician attempted to re-sheath the subject flow diverter but increased resistance was felt when the subject flow diverter was re-sheathed.The subject flow diverter along with the microcatheter were removed from the patients vasculature.The subject flow diverter and the microcatheter were replaced and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that a neurovascular surgical procedure was performed in a patient with cavernous aneurysm.During the procedure the physician felt resistance while advancing the subject flow diverter to the target lesion.At the target lesion the physician felt the subject flow diverter didn¿t feel as smooth as usual and didn¿t want to force it to open, therefore the physician attempted to re-sheath the subject flow diverter but increased resistance was felt when the subject flow diverter was re-sheathed.The subject flow diverter along with the microcatheter were removed from the patients vasculature.The subject flow diverter and the microcatheter were replaced and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
D4 expiration date - added.H4 manufacturing date ¿ added.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that the device may not have been prepped correctly.After the case it was questioned if the tech had prepped the device properly.Further details were not provided.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to the as reported 'flow diverter stent difficult/unable to capture into micro-catheter', 'stent difficult/unable to advance or pullback through catheter' for this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 17MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14114587
Report Number3008881809-2022-00178
Device Sequence Number13983859
Product Code OUT
UDI-Device Identifier07613327386530
UDI-Public07613327386530
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2024
Device Model NumberFDS40017
Device Catalogue NumberFDS40017
Device Lot Number22939063
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CATALYST 5 CATHETER (STRYKER); EXCELSIOR XT-27 MICROCATHETER (STRYKER)