| Model Number |
B2-70071-DF |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zyno medical is waiting for the arrival of the affected device.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, zyno medical received a complaint from an end user that "tubing connected to the patient, heard a loud pop during the infusion and it was leaking from the filter.Checked clamps, and none were engaged.Switched out tubing, and it worked without issue on the next tubing." a patient was involved but was not harmed or injured.The device operator was a registered nurse.The medication being infused was remicade.The contract manufacturer of the affected device is becton, dickinson and company.
|
| |
|
Additional Manufacturer Narrative
|
|
The service provider provided the evaluation report on 06/22/2022.The actual device was tested.Upon priming the set, leaking was observed from the right side of the 0.2-micron filter.The reported issue was confirmed.The device does not meet the passing criteria.The device will be returned to the contract manufacturer for further evaluation.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2022-00012).
|
| |
|