WRIGHTS LANE SYNTHES USA PRODUCTS LLC INSERTION HANDLE/ RADIOLUCENT MEDIUM; SURGICAL IMPLANT/TRIAL-IMPLANT/SIZER HOLDER, REUSABLE
|
Back to Search Results |
|
| Model Number |
03.043.021 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This is report 1 of 3 for (b)(4).It was reported that during an unknown surgery on (b)(6) 2022, it was observed that the when the doctor was drilling the proximal medial to make a lateral static hole in an ans tibia nail, he noticed that the drill bit was making contact with the implant.It was further reported that the driving cap was also stuck in the nail insertion handle.The procedure was completed successfully without delay.There were no adverse patient consequences reported.No additional information was provided.This report is for one (1) insertion handle/ radiolucent medium.
|
| |
|
Additional Manufacturer Narrative
|
|
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Udi: (b)(4).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: part number: 03.043.021.Lot number: 81p3383.Manufacturing site: haegendorf.Release to warehouse date: january 26, 2021.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|