• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Arteriosclerosis/ Atherosclerosis (4437); Restenosis (4576)
Date of Event 03/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient identifier: (b)(6).Age at time of event: (b)(6).
 
Event or Problem Description
(b)(4) clinical trial: it was reported this stent had restenosis and internal device fracture.The subject underwent treatment with the 6.0mm x 100mm eluvia drug eluting stent on (b)(6) 2021 as a part of the elegance clinical trial.The target lesion was located in the right distal superficial femoral artery (sfa) with proximal reference vessel diameter of 5mm and distal reference vessel diameter of 5mm.The lesion was 100mm long , 100% stenosed, and classified as tasc ii b lesion.Prior to target lesion treatment, pre dilation was performed using a 5mm x 150mm mustang percutaneous transluminal angioplasty (pta) balloon.Treatment of the target lesion was performed by implanting a 6.0mm x 100mm eluvia, and the residual stenosis was noted to be 0%.The patient was discharged the same day on clopidogrel.During treatment of the target lesion, dissection of grade b was noted post angioplasty with 5 mm x 150 mm mustang balloon.In response to the event 6.0 mm x 100 mm eluvia stent was implanted.Following post dilatation with 5.0 mm x 150 mm mustang pta balloon, the final residual stenosis was noted to be 0%.The complication was resolved.On (b)(6) 2022, the subject presented with the symptoms of worsened pain of bilateral calves (right leg worse than left).On the same day, ankle brachial indices at rest were measured which revealed right 0.61 and left 0.86.Toe brachial indices at rest revealed right 0.58 and left 0.81, and absolute great toe pressure at rest were right 81 mmhg and left 113 mmhg.The ankle pulse volume recording waveforms were of moderately abnormal amplitude on the right and mildly abnormal amplitude with bunted dicrotic notch on the left.There was dampened flow in all five right five digits and dampened flow in all the left digits by photoplethysmography.An additional vascular evaluation revealed monophasic doppler signal of right posterior tibial and peroneal artery; no wounds or ulcers were noted.On (b)(6) 2022, the subject revisited the hospital for planned lower extremity arteriogram which revealed the right sfa was occluded from the ostium.The stent was occluded and fractured.No additional intervention had been performed at this time, however bypass surgery was recommended at a later date.
 
Additional Manufacturer Narrative
A1 - patient identifier: (b)(6).A2 - age at time of event: 66 years.
 
Event or Problem Description
Elegance clinical trial it was reported this stent had restenosis and internal device fracture.The subject underwent treatment with the 6.0mm x 100mm eluvia drug eluting stent on (b)(6) 2021 as a part of the elegance clinical trial.The target lesion was located in the right distal superficial femoral artery (sfa) with proximal reference vessel diameter of 5mm and distal reference vessel diameter of 5mm.The lesion was 100mm long , 100% stenosed, and classified as tasc ii b lesion.Prior to target lesion treatment, pre dilation was performed using a 5mm x 150mm mustang percutaneous transluminal angioplasty (pta) balloon.Treatment of the target lesion was performed by implanting a 6.0mm x 100mm eluvia, and the residual stenosis was noted to be 0%.The patient was discharged the same day on clopidogrel.During treatment of the target lesion, dissection of grade b was noted post angioplasty with 5 mm x 150 mm mustang balloon.In response to the event 6.0 mm x 100 mm eluvia stent was implanted.Following post dilatation with 5.0 mm x 150 mm mustang pta balloon, the final residual stenosis was noted to be 0%.The complication was resolved.On (b)(6) 2022, the subject presented with the symptoms of worsened pain of bilateral calves (right leg worse than left).On the same day, ankle brachial indices at rest were measured which revealed right 0.61 and left 0.86.Toe brachial indices at rest revealed right 0.58 and left 0.81, and absolute great toe pressure at rest were right 81 mmhg and left 113 mmhg.The ankle pulse volume recording waveforms were of moderately abnormal amplitude on the right and mildly abnormal amplitude with bunted dicrotic notch on the left.There was dampened flow in all five right five digits and dampened flow in all the left digits by photoplethysmography.An additional vascular evaluation revealed monophasic doppler signal of right posterior tibial and peroneal artery; no wounds or ulcers were noted.On (b)(6) 2022, the subject revisited the hospital for planned lower extremity arteriogram which revealed the right sfa was occluded from the ostium.The stent was occluded and fractured.No additional intervention had been performed at this time, however bypass surgery was recommended at a later date.It was further reported that on (b)(6) 2022, the subject revisited the hospital for the planned bypass surgery and was hospitalized on the same day for further medical intervention due to target lesion restenosis.The lesion was treated by performing right femoral to popliteal artery bypass surgery.
 
Additional Manufacturer Narrative
A1 - patient identifier: (b)(4).A2 - age at time of event: patient was 66 years at time of study enrollment.
 
Event or Problem Description
Elegance clinical trial it was reported this stent had restenosis and internal device fracture.The subject underwent treatment with the 6.0mm x 100mm eluvia drug eluting stent on (b)(6) 2021 as a part of the elegance clinical trial.The target lesion was located in the right distal superficial artery with proximal reference vessel diameter of 5mm and distal reference vessel diameter of 5mm.The lesion was 100mm long , 100% stenosed, and classified as tasc ii b lesion.Prior to target lesion treatment, pre dilation was performed using a 5mm x 150mm mustang percutaneous transluminal angioplasty (pta) balloon.Treatment of the target lesion was performed by implanting a 6.0mm x 100mm eluvia, and the residual stenosis was noted to be 0%.The patient was discharged the same day on clopidogrel.On (b)(6) 2022, the subject presented with the symptoms of worsened pain of bilateral calves (right leg worse than left).On the same day, ankle brachial indices at rest were measured which revealed right 0.61 and left 0.86.Toe brachial indices at rest revealed right 0.58 and left 0.81, and absolute great toe pressure at rest were right 81 mmhg and left 113 mmhg.The ankle pulse volume recording waveforms were of moderately abnormal amplitude on the right and mildly abnormal amplitude with bunted dicrotic notch on the left.There was dampened flow in all five right five digits and dampened flow in all the left digits by photoplethysmography.An additional vascular evaluation revealed monophasic doppler signal of right posterior tibial and peroneal artery; no wounds or ulcers were noted.On (b)(6) 2022, the subject revisited the hospital for planned lower extremity arteriogram which revealed the right common iliac artery had mild non-flow limiting stenosis, the right anterior tibial artery was diffusely diseased and diminutive in the distal calf, the right tibioperoneal trunk artery had mild to moderate stenosis, and the right superficial femoral artery (sfa) was occluded from ostium.The site reported a stent fracture and occlusion of the study device.No additional intervention had been performed at this time, however bypass surgery was recommended at a later date.It was further reported that on (b)(6) 2022, the subject revisited the hospital for the planned bypass surgery and was hospitalized on the same day for further medical intervention due to target lesion restenosis.The lesion was treated by performing right femoral to popliteal artery bypass surgery.If was further reported that on the same day of index procedure, 70% stenosis noted in the right external iliac artery was treated with 7 mmx 40 mm balloon followed by placement of 8 mm x 40 mm self-expanding stent.Subsequently, high-grade stenosis noted in the left external iliac artery was treated with 7 mm x 40 mm balloon and 8 mm x 40 mm self-expanding stent.Post treatment, no residual stenosis was noted.On (b)(6) 2022, subject visited the hospital as follow-up of the symptoms of claudication that occurred after ambulating short distance and was lifestyle limiting.Subject continued to have right leg pain worse than the left leg.No evidence of rest pain or wounds/ulcers were noted, and rutherford category was assessed as 3.Subject was also noted with soreness in the hips which was likely musculoskeletal in nature and did not align with the blockage found on angiogram.Based on the findings from the study results and arteriogram, subject was recommended to undergo right femoral to popliteal graft in mid-april.Subject was instructed to ambulate 5 days a week and to continue aspirin.On (b)(6) 2022, subject was scheduled for right femoral to above knee popliteal bypass surgery, however, subject had developed groin folliculitis bilaterally and the procedure was not performed.Subject was sent home with doxycycline and the right femoral to above knee popliteal bypass surgery was rescheduled on (b)(6) 2022.On (b)(6) 2022, the subject visited the hospital for the planned bypass surgery and was hospitalized on the same day for further medical intervention.On (b)(6) 2022, 185 days post index procedure, 100% stenosis noted in the superficial femoral artery was treated by performing right femoral to above knee popliteal artery bypass surgery with 8 mm fusion bioline graft.In addition, endarterectomy of common femoral artery was performed using freer elevator during the procedure.On (b)(6) 2022, the event was considered resolved and on the same day, the subject was discharged from the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14115904
Report Number2134265-2022-03571
Device Sequence Number5932888
Product Code NIU
UDI-Device Identifier08714729876601
UDI-Public08714729876601
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P180011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2023
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0027593574
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer04/26/2022
05/26/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date05/19/2022
06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient SexMale
Patient RaceWhite
-
-