• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL PE ADULT-PED WET LF 6/CS; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE ADULT-PED WET LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028446
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history record of batch number 74m2101787 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.
 
Event or Problem Description
Air heard hissing from pleurevac.Air leak noted.Entire pleurevac changed and then no issue.No visible crack but air leak from red/blue connector.The same issue has happened with all of these chest drains.2 of the 6 were used in a clinical setting.The remaining drains were tested outside of using on a patient and the customer found similar issues on them.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED WET LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key14116024
Report Number3004365956-2022-00030
Device Sequence Number14189340
Product Code KDQ
UDI-Device Identifier34026704631279
UDI-Public34026704631279
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/16/2024
Device Model NumberIPN028446
Device Catalogue NumberA-7000-08LF
Device Lot Number74M2101787
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-