• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL UNIVERSAL ROD, SHORT T2 HUMERUS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL UNIVERSAL ROD, SHORT T2 HUMERUS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 1806-0113
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "universal rod broke at the connection when extracting a t2 humeral nail that was inserted about 10 years ago.Because a new fracture occurred distal to the nail, it was decided to treat with a plate, and the nail was extracted this time.The nail could not be removed due to breakage of the universal rod, so another attempt is made to remove the nail on the (b)(6) 2022.".
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
 
Event or Problem Description
As reported: "universal rod broke at the connection when extracting a t2 humeral nail that was inserted about 10 years ago.Because a new fracture occurred distal to the nail, it was decided to treat with a plate, and the nail was extracted this time.The nail could not be removed due to breakage of the universal rod, so another attempt is made to remove the nail on the (b)(6) 2022.".
 
Additional Manufacturer Narrative
Please note corrections to d9/h3 and h6 method code as the device was returned for evaluation.The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the returned universal rod shows signs of intense use evident by the appearance of scratches and minor dents all over the surface of the rod.The tip of the rod was found to be broken.The nature of the breakage indicates towards a brittle fracture due to a sudden impact load.The outer end face of the handle has dents indicating extensive hitting.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a user related issue.The nature of breakage indicates towards a brittle fracture caused most likely caused due to high impact forces during nail extraction.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL ROD, SHORT T2 HUMERUS
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14116135
Report Number0009610622-2022-00138
Device Sequence Number4335065
Product Code LXH
UDI-Device Identifier04546540196026
UDI-Public04546540196026
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1806-0113
Device Catalogue Number18060113
Device Lot NumberK730100
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2003
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-