| Model Number |
INSPIRATORY LINE HTD 8FT ADULT 20/CS |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the 4681-504-inspiratory line htd 8ft adult 20/cs started to burned while connected to a patient.The customer confirm that there is no harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation- the sample was visually inspected and it was found that the wire is burned and the wire does not contain the terminal connector with the high current connector, hence the sample was received incomplete.The reported issue was confirmed.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|