MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE HIGH; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3542711 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Crushing Injury (1797)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old female patient underwent a breast augmentation procedure with a mentor siltex contour profile high 275cc saline breast prosthesis that deflated post implantation.Deflation of the patient¿s right breast prosthesis was diagnosed via a physical exam.Additionally, it was reported that the deflation may have been caused by a mammogram.As a result, the patient underwent bilateral explantation with no replacement breast prostheses on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis deflation.Manufacturer¿s reference number: pc-(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 27-apr-2022, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a deflation in the breast implant after a mammogram.The product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and a tear was observed on the posterior view, measuring approximately 0.8 cm.The evaluation determined that the cause of the deflation is the trauma experienced.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of the saline-filled mammary prosthesis.It should be noted that as part of the mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|