• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. HEIGHT EXPANSION DRIVER; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. HEIGHT EXPANSION DRIVER; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 2112-90007
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that outside the or the tips of two capri height expansion drivers looked worn.The reported event was noted during inspection.There is no procedure associated with this reported event, no patient involvement, and no adverse consequence has been reported.This report captures the second height expansion driver.
 
Additional Manufacturer Narrative
A visual inspection was performed and inspection confirmed the driver tips to be worn as reported by the customer.All driver teeth deformed in a groove-like manner from repeated rotation and abrasion about the driver tip, relatively uniform degree of deformation across all driver teeth and deformation located where screw makes contact with driver when engaged.Device history records were reviewed for this lot, and no relevant manufacturing issues were identified.Complaint history records for reviewed for this lot number, similar complaints were identified.The observed deformation and age of the device (6 years) indicate the failure was likely caused by normal wear.
 
Event or Problem Description
It was reported that outside the or the tips of two capri height expansion drivers looked worn.The reported event was noted during inspection.There is no procedure associated with this reported event, no patient involvement, and no adverse consequence has been reported.This report captures the second height expansion driver.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEIGHT EXPANSION DRIVER
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14116265
Report Number3004774118-2022-00154
Device Sequence Number2428134
Product Code MQP
UDI-Device Identifier10888857033641
UDI-Public10888857033641
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K142016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2112-90007
Device Catalogue Number2112-90007
Device Lot NumberELDG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-