| Model Number |
105-7100-060 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Improper Chemical Reaction (2952)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that an onyx had poor visualization because the onyx did not mix properly.The patient was being treated for a arteriovenous malformation (avm) grade 3 that was not in a n eloquent region.Vessel tortuosity was moderate.Angiographic result post procedure was good.The devices were prepared according to the instructions for use (ifu). onyx 18 had been shaking at 8:30am at speed of 8 for 30 minutes was then brought down to speed of 5 and drawn up to inject at 10:45am.Physician injected onyx into microcatheter to treat avm and said onyx was not visible under fluoro because it was not shaken properly.The onyx was shaken for 2 hours.It did not sit for more than 5 minutes prior to injection.The physician did not heat, shake, and re-heat the onyx vials prior to aspiration.The physician injected the onyx immediately after mixing. physician used onyx 34 prior with no problems.Case with completed.No patient symptoms were reported. ancillary devices: headway duo 156 microcatheter, synchro 10 guidewire.
|
| |
|