AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number ZCT150
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Weight and ethnicity: unknown, as information was asked but not provided.Date of event: unknown as information was asked but not provided.Date implanted: not applicable, as the iol was removed and replaced during the same procedure.Date explanted: not applicable, as the iol was removed and replaced during the same procedure.The device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device/lot history record and historical data analysis for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported the intraocular lens (iol) was inserted into the patient's ocular dexter (right eye) and a scratch on the lens was identified after lens implantation.The iol was removed during the same procedure.Through follow-up, additional information was received the same procedure was completed successfully using back-up lens with the same model and diopter size.There was no unplanned vitrectomy.Patient outcome post-procedure was reported as unknown.No further information is available.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation: yes.Section d-9: date returned to manufacturer: 22-apr-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: the suspect iol was received stuck to the patient label.The lens was removed and found to be cut in half.Visual inspection under magnification was performed and revealed scratches and an edge chip on the lens.No further issues were identified.Complaint issue "cosmetic issues" was observed during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14116499
Report Number3012236936-2022-00900
Device Sequence Number16733206
Product Code MJP
UDI-Device Identifier05050474552227
UDI-Public(01)05050474552227(17)220903
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2022
Device Model NumberZCT150
Device Catalogue NumberZCT150U240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer06/14/2022
Initial Report FDA Received Date04/14/2022
Supplement Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-