• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2022
Event Type  Injury  
Event Description
Information was received from a healthcare professional (hcp) via a company representative (rep) regarding a patient receiving morphine (50 mg/ml at 6 mg/day) via an implanted pump. The indication for pump use was unknown. It was reported that the patient¿s catheter was not patent. The hcp was unable to aspirate cerebrospinal fluid (csf) during a dye study. The catheter was replaced on (b)(6) 2022 and discarded. There were no environmental, external, or patient factors that may have led or contributed to the issue. The issue was resolved at the time of this report and the patient¿s status was ¿alive- no injury¿. The patient¿s weight and medical history were asked but unknown.
 
Manufacturer Narrative
Concomitant medical products: product id: 8780; serial#: (b)(4); implanted: (b)(6) 2016; explanted: (b)(6) 2022; product type: catheter. Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 28-jul-2018, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14116583
MDR Text Key290568443
Report Number3004209178-2022-04779
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/28/2018
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured10/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/14/2022 Patient Sequence Number: 1
-
-