ST. JUDE MEDICAL PEEL-AWAY INTRODUCER 14 CM SHEATH INTRODUCER KIT, 7F; INTRODUCER, CATHETER Back to Search Results
Model Number 405108
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
During icd implant procedure, an axillary vein was punctured and at the time of passing the guidewire it did not pass through the needle and was stuck.Another angiographic needle and another short guidewire were used, the same peel away was used.A new puncture of the left subclavian vein is required due to inadequate functioning of the needle and the guidewire, with the guidewire getting stuck in the lumen of the needle.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported insertion difficulties and subsequent device replacement remain unknown.
 
Additional Manufacturer Narrative
One 7f peel-away introducer guidewire and needle with cover were received for evaluation.The proximal end of the guidewire was pulled to remove the guidewire from the needle; resistance was noted due to a reddish/black foreign material (fm) attached to the guidewire surface, however, the guidewire was able to be removed.Additional testing revealed the returned fm appeared to match generic biological material.The guidewire outside diameter and needle inside diameter measurements were consistent with manufacturing specifications.The guidewire was able to be inserted fully through the returned needle after the fm was removed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the guidewire stuck is consistent with the biological material on the guidewire.The cause of the biological material on the guidewire remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEEL-AWAY INTRODUCER 14 CM SHEATH INTRODUCER KIT, 7F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14116741
Report Number3005334138-2022-00260
Device Sequence Number14026051
Product Code DYB
UDI-Device Identifier05414734204101
UDI-Public05414734204101
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K894431
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number405108
Device Catalogue Number405108
Device Lot Number8203383
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer05/17/2022
06/28/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date05/23/2022
07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexMale
Patient Weight70 KG
-
-